Development and study of characteristics of stripped detectors on the basis of CVD-diamond
The prototypes of the coordinate-sensitive semiconductor detectors based on polycrystalline diamond films (CVD-diamond) have been manufactured in the NSC KIPT. In this communication, a version of the detector with the strip-type contacts and its characteristics are described. The techniques of the...
Збережено в:
Дата: | 2016 |
---|---|
Автори: | , , , , , , , , , , |
Формат: | Стаття |
Мова: | English |
Опубліковано: |
Національний науковий центр «Харківський фізико-технічний інститут» НАН України
2016
|
Назва видання: | Вопросы атомной науки и техники |
Теми: | |
Онлайн доступ: | http://dspace.nbuv.gov.ua/handle/123456789/115391 |
Теги: |
Додати тег
Немає тегів, Будьте першим, хто поставить тег для цього запису!
|
Назва журналу: | Digital Library of Periodicals of National Academy of Sciences of Ukraine |
Цитувати: | Development and study of characteristics of stripped detectors on the basis of CVD-diamond / V.E. Kutny, A.A. Vierovkin, A.V. Rybka, S.F. Dudnik, O.A. Opalev, V.E. Strelnitsky, A.A. Zakharchenko, A.Eh. Tenishev, V.L. Uvarov, V.A. Shevchenko, I.N. Shlyakhov // Вопросы атомной науки и техники. — 2016. — № 3. — С. 116-119. — Бібліогр.: 2 назв. — англ. |
Репозитарії
Digital Library of Periodicals of National Academy of Sciences of UkraineРезюме: | The prototypes of the coordinate-sensitive semiconductor detectors based on polycrystalline diamond films
(CVD-diamond) have been manufactured in the NSC KIPT. In this communication, a version of the detector with
the strip-type contacts and its characteristics are described. The techniques of the detector calibration both as a
whole and for separate groups of the strips against the absorbed dose rate of the electron and bremsstrahlung radiation
are developed. By means of the computer simulations, an independent analysis of conditions of the radiation
field registration with the detector has been conducted. The experiments carried out at the LU-10 accelerator at NSC
KIPT have shown that the sensitivities of the separate elements (strips) of the detector are similar. The results of the
measurements of the volume sensitivity of the detector are in agreement with the data received earlier for a singlecontact
detector. The possibility of the detector application in the study of transverse dynamics of an electron beam
is shown. |
---|