Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination

The characteristics of conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination have been studied. Urease has been immobilized by cross-linking with bovine serum albumin using glutaraldehyde. The resulting conductivity changes are produced by enzymatic...

Повний опис

Збережено в:
Бібліографічні деталі
Дата:1996
Автори: Dzyadevich, S.V., Zhylyak, G.A., Soldatkin, A.P.
Формат: Стаття
Мова:English
Опубліковано: Інститут молекулярної біології і генетики НАН України 1996
Назва видання:Биополимеры и клетка
Онлайн доступ:http://dspace.nbuv.gov.ua/handle/123456789/153663
Теги: Додати тег
Немає тегів, Будьте першим, хто поставить тег для цього запису!
Назва журналу:Digital Library of Periodicals of National Academy of Sciences of Ukraine
Цитувати:Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination / S.V. Dzyadevich, G.A. Zhylyak, A.P. Soldatkin // Биополимеры и клетка. — 1996. — Т. 12, № 1. — С. 53-57. — Бібліогр.: 9 назв. — англ.

Репозитарії

Digital Library of Periodicals of National Academy of Sciences of Ukraine
id irk-123456789-153663
record_format dspace
spelling irk-123456789-1536632019-07-04T20:44:58Z Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination Dzyadevich, S.V. Zhylyak, G.A. Soldatkin, A.P. The characteristics of conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination have been studied. Urease has been immobilized by cross-linking with bovine serum albumin using glutaraldehyde. The resulting conductivity changes are produced by enzymatically catalyzed hydrolysis of urea. This process for both immobilized enzyme and soluble one is described by the classic laws of enzymatic kinetics. The influence of ionic strength, pH and buffer capacity of the samples on the biosensor response has been thoroughly tested. The results have been discussed concerning urea concentration analysis in human blood. 1996 Article Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination / S.V. Dzyadevich, G.A. Zhylyak, A.P. Soldatkin // Биополимеры и клетка. — 1996. — Т. 12, № 1. — С. 53-57. — Бібліогр.: 9 назв. — англ. 0233-7657 http://dspace.nbuv.gov.ua/handle/123456789/153663 577.15+573.6 en Биополимеры и клетка Інститут молекулярної біології і генетики НАН України
institution Digital Library of Periodicals of National Academy of Sciences of Ukraine
collection DSpace DC
language English
description The characteristics of conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination have been studied. Urease has been immobilized by cross-linking with bovine serum albumin using glutaraldehyde. The resulting conductivity changes are produced by enzymatically catalyzed hydrolysis of urea. This process for both immobilized enzyme and soluble one is described by the classic laws of enzymatic kinetics. The influence of ionic strength, pH and buffer capacity of the samples on the biosensor response has been thoroughly tested. The results have been discussed concerning urea concentration analysis in human blood.
format Article
author Dzyadevich, S.V.
Zhylyak, G.A.
Soldatkin, A.P.
spellingShingle Dzyadevich, S.V.
Zhylyak, G.A.
Soldatkin, A.P.
Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination
Биополимеры и клетка
author_facet Dzyadevich, S.V.
Zhylyak, G.A.
Soldatkin, A.P.
author_sort Dzyadevich, S.V.
title Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination
title_short Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination
title_full Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination
title_fullStr Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination
title_full_unstemmed Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination
title_sort conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination
publisher Інститут молекулярної біології і генетики НАН України
publishDate 1996
url http://dspace.nbuv.gov.ua/handle/123456789/153663
citation_txt Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination / S.V. Dzyadevich, G.A. Zhylyak, A.P. Soldatkin // Биополимеры и клетка. — 1996. — Т. 12, № 1. — С. 53-57. — Бібліогр.: 9 назв. — англ.
series Биополимеры и клетка
work_keys_str_mv AT dzyadevichsv conductometricureasemicrobiosensorbasedonthinfilminterdigitatedelectrodesforureadetermination
AT zhylyakga conductometricureasemicrobiosensorbasedonthinfilminterdigitatedelectrodesforureadetermination
AT soldatkinap conductometricureasemicrobiosensorbasedonthinfilminterdigitatedelectrodesforureadetermination
first_indexed 2023-05-20T17:42:15Z
last_indexed 2023-05-20T17:42:15Z
_version_ 1796153903022080000