Импедансная модель прецизионных платиновых термометров сопротивления в области высоких температур

Double-pole multiunit impedance model of sensitive thermometer element during measurement on alternating current is proposed. It takes into consideration impact on resistance of sensitive element of bypassing effect which is caused by essential lowing of specific resistance of framework isolation el...

Повний опис

Збережено в:
Бібліографічні деталі
Дата:2012
Автори: Михаль, А.А., Мелещук, Д.В.
Формат: Стаття
Мова:Russian
Опубліковано: Інститут електродинаміки НАН України 2012
Назва видання:Технічна електродинаміка
Теми:
Онлайн доступ:http://dspace.nbuv.gov.ua/handle/123456789/62076
Теги: Додати тег
Немає тегів, Будьте першим, хто поставить тег для цього запису!
Назва журналу:Digital Library of Periodicals of National Academy of Sciences of Ukraine
Цитувати:Импедансная модель прецизионных платиновых термометров сопротивления в области высоких температур / А.А. Михаль, Д.В. Мелещук // Технічна електродинаміка. — 2012. — № 4. — С. 73–79. — Бібліогр.: 10 назв. — pос.

Репозитарії

Digital Library of Periodicals of National Academy of Sciences of Ukraine
Опис
Резюме:Double-pole multiunit impedance model of sensitive thermometer element during measurement on alternating current is proposed. It takes into consideration impact on resistance of sensitive element of bypassing effect which is caused by essential lowing of specific resistance of framework isolation elements with temperature growth. As a result, during the measurement of thermometer resistance on direct and alternating current the error appears which is the function of measurable temperature and current frequency. Theoretical investigations of frequency error of thermometers in the range of temperatures from 0 to 1100 °C are conducted. Difference in the results of measurement of resistance of major types of thermometers is established which leads to the error of 0,1-0,2 mK. Model analysis demonstrates the necessity for introduction of individual frequency correction to the measurement result like correction for zero power of dispersion.