Comparison of two data acquisition and processing systems of Moller polarimeter in Hall A of Jefferson Lab
Two data acquisition and processing systems are used simultaneously to measure electron beam polarization by Moller polarimeter in Hall A of Jefferson Lab (Newport News, VA, USA). The old system (since 1997) is fully functional, but is not repairable in case of malfunction (system modules are not...
Saved in:
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Національний науковий центр «Харківський фізико-технічний інститут» НАН України
2014
|
Series: | Вопросы атомной науки и техники |
Subjects: | |
Online Access: | http://dspace.nbuv.gov.ua/handle/123456789/80275 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Two data acquisition and processing systems are used simultaneously to measure electron beam polarization by
Moller polarimeter in Hall A of Jefferson Lab (Newport News, VA, USA). The old system (since 1997) is fully
functional, but is not repairable in case of malfunction (system modules are not manufactured anymore). The new
system (since 2010) based on flash-ADC is more accurate, but currently requires more detailed adjustment and further
improvement. Description and specifications of two data acquisition and processing systems have been given.
The results of polarization measurements during experiments conducted in Hall A from 2010 to 2012 are compared. |
---|