Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields

Results of a study of dosimetry characteristics of a prototype of the detector based on a polycrystalline diamond film (CVD-diamond) produced in NSC KIPT are summarized. The techniques of the detector calibration against the electron and X-ray radiation dose rate are developed. The conditions of c...

Повний опис

Збережено в:
Бібліографічні деталі
Дата:2014
Автори: Kutny, V.E., Nikiforov, V.I., Opalev, O.A., Rogov, Yu.V., Rybka, A.V., Shevchenko, V.A., Shlyakhov, I.N., Strelnitsky, V.E., Tenishev, A.Eh., Uvarov, V.L.
Формат: Стаття
Мова:English
Опубліковано: Національний науковий центр «Харківський фізико-технічний інститут» НАН України 2014
Назва видання:Вопросы атомной науки и техники
Теми:
Онлайн доступ:http://dspace.nbuv.gov.ua/handle/123456789/80276
Теги: Додати тег
Немає тегів, Будьте першим, хто поставить тег для цього запису!
Назва журналу:Digital Library of Periodicals of National Academy of Sciences of Ukraine
Цитувати:Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields / V.E. Kutny, V.I. Nikiforov, O.A. Opalev, Yu.V. Rogov, A.V. Rybka, V.A. Shevchenko, I.N. Shlyakhov, V.E. Strelnitsky, A.Eh. Tenishev, V.L. Uvarov // Вопросы атомной науки и техники. — 2014. — № 3. — С. 162-165. — Бібліогр.: 12 назв. — англ.

Репозитарії

Digital Library of Periodicals of National Academy of Sciences of Ukraine
Опис
Резюме:Results of a study of dosimetry characteristics of a prototype of the detector based on a polycrystalline diamond film (CVD-diamond) produced in NSC KIPT are summarized. The techniques of the detector calibration against the electron and X-ray radiation dose rate are developed. The conditions of calibration were studied by means of computer simulation. For determination of detector sensitivity to electron radiation, it was placed inside of a standard polystyrene phantom. An additional filter of electrons was used at measurement in the X-ray field. Every time detector irradiation was carried out together with the Harwell Red Perpex 4034 dosimeters, which registrations provided the calibration data. The measurements, executed at the LU-10, EPOS and LU-40 linacs of NSC KIPT, have demonstrated, that the values of detector sensitivity against each type of radiation are close and conform to ones obtained earlier for the low-intensity radiation fields. Considering significant radiation durability of the CVD-diamond, it enables its application in technological dosimetry.