Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields
Results of a study of dosimetry characteristics of a prototype of the detector based on a polycrystalline diamond film (CVD-diamond) produced in NSC KIPT are summarized. The techniques of the detector calibration against the electron and X-ray radiation dose rate are developed. The conditions of c...
Збережено в:
Дата: | 2014 |
---|---|
Автори: | , , , , , , , , , |
Формат: | Стаття |
Мова: | English |
Опубліковано: |
Національний науковий центр «Харківський фізико-технічний інститут» НАН України
2014
|
Назва видання: | Вопросы атомной науки и техники |
Теми: | |
Онлайн доступ: | http://dspace.nbuv.gov.ua/handle/123456789/80276 |
Теги: |
Додати тег
Немає тегів, Будьте першим, хто поставить тег для цього запису!
|
Назва журналу: | Digital Library of Periodicals of National Academy of Sciences of Ukraine |
Цитувати: | Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields / V.E. Kutny, V.I. Nikiforov, O.A. Opalev, Yu.V. Rogov, A.V. Rybka, V.A. Shevchenko, I.N. Shlyakhov, V.E. Strelnitsky, A.Eh. Tenishev, V.L. Uvarov // Вопросы атомной науки и техники. — 2014. — № 3. — С. 162-165. — Бібліогр.: 12 назв. — англ. |
Репозитарії
Digital Library of Periodicals of National Academy of Sciences of UkraineРезюме: | Results of a study of dosimetry characteristics of a prototype of the detector based on a polycrystalline diamond
film (CVD-diamond) produced in NSC KIPT are summarized. The techniques of the detector calibration against the
electron and X-ray radiation dose rate are developed. The conditions of calibration were studied by means of computer
simulation. For determination of detector sensitivity to electron radiation, it was placed inside of a standard
polystyrene phantom. An additional filter of electrons was used at measurement in the X-ray field. Every time detector
irradiation was carried out together with the Harwell Red Perpex 4034 dosimeters, which registrations provided
the calibration data. The measurements, executed at the LU-10, EPOS and LU-40 linacs of NSC KIPT, have demonstrated,
that the values of detector sensitivity against each type of radiation are close and conform to ones obtained
earlier for the low-intensity radiation fields. Considering significant radiation durability of the CVD-diamond, it
enables its application in technological dosimetry. |
---|