Detectors and instruments for detection of X-ray radiation on the basis of zinc selenide crystals

Theoretical analysis and experimental studies have been carried out of parameters of two-layer sandwich-detector of scintillator-photodiode type, in which the low-energy detector works both as absorption filter of low-energy fraction of radiation for the high-energy detector and detector of the low-...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Functional Materials
Datum:2015
Hauptverfasser: Ryzhikov, V.D., Volkov, V.G., Opolonin, O.D., Makhota, S.V., Maksymchuk, V.V.
Format: Artikel
Sprache:English
Veröffentlicht: НТК «Інститут монокристалів» НАН України 2015
Schlagworte:
Online Zugang:https://nasplib.isofts.kiev.ua/handle/123456789/119702
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Назва журналу:Digital Library of Periodicals of National Academy of Sciences of Ukraine
Zitieren:Detectors and instruments for detection of X-ray radiation on the basis of zinc selenide crystals / V.D. Ryzhikov, V.G. Volkov, O.D. Opolonin, S.V. Makhota, V.V. Maksymchuk // Functional Materials. — 2015. — Т. 22, № 4. — С. 552-558. — Бібліогр.: 18 назв. — англ.

Institution

Digital Library of Periodicals of National Academy of Sciences of Ukraine
id nasplib_isofts_kiev_ua-123456789-119702
record_format dspace
spelling Ryzhikov, V.D.
Volkov, V.G.
Opolonin, O.D.
Makhota, S.V.
Maksymchuk, V.V.
2017-06-08T07:08:02Z
2017-06-08T07:08:02Z
2015
Detectors and instruments for detection of X-ray radiation on the basis of zinc selenide crystals / V.D. Ryzhikov, V.G. Volkov, O.D. Opolonin, S.V. Makhota, V.V. Maksymchuk // Functional Materials. — 2015. — Т. 22, № 4. — С. 552-558. — Бібліогр.: 18 назв. — англ.
1027-5495
DOI: dx.doi.org/10.15407/fm22.04.552
https://nasplib.isofts.kiev.ua/handle/123456789/119702
Theoretical analysis and experimental studies have been carried out of parameters of two-layer sandwich-detector of scintillator-photodiode type, in which the low-energy detector works both as absorption filter of low-energy fraction of radiation for the high-energy detector and detector of the low-energy radiation. The energy dependence deviation does not exceed ±5 % in the range from 30 to 120 keV. The developed detector and the instrument designed on its base allow contactless measurement of anode voltage of X-ray emitter from 40 to 140 kV with accuracy of ±3 %. The dose rate measurement range is 1-200 R/min.
en
НТК «Інститут монокристалів» НАН України
Functional Materials
Devices and instruments
Detectors and instruments for detection of X-ray radiation on the basis of zinc selenide crystals
Article
published earlier
institution Digital Library of Periodicals of National Academy of Sciences of Ukraine
collection DSpace DC
title Detectors and instruments for detection of X-ray radiation on the basis of zinc selenide crystals
spellingShingle Detectors and instruments for detection of X-ray radiation on the basis of zinc selenide crystals
Ryzhikov, V.D.
Volkov, V.G.
Opolonin, O.D.
Makhota, S.V.
Maksymchuk, V.V.
Devices and instruments
title_short Detectors and instruments for detection of X-ray radiation on the basis of zinc selenide crystals
title_full Detectors and instruments for detection of X-ray radiation on the basis of zinc selenide crystals
title_fullStr Detectors and instruments for detection of X-ray radiation on the basis of zinc selenide crystals
title_full_unstemmed Detectors and instruments for detection of X-ray radiation on the basis of zinc selenide crystals
title_sort detectors and instruments for detection of x-ray radiation on the basis of zinc selenide crystals
author Ryzhikov, V.D.
Volkov, V.G.
Opolonin, O.D.
Makhota, S.V.
Maksymchuk, V.V.
author_facet Ryzhikov, V.D.
Volkov, V.G.
Opolonin, O.D.
Makhota, S.V.
Maksymchuk, V.V.
topic Devices and instruments
topic_facet Devices and instruments
publishDate 2015
language English
container_title Functional Materials
publisher НТК «Інститут монокристалів» НАН України
format Article
description Theoretical analysis and experimental studies have been carried out of parameters of two-layer sandwich-detector of scintillator-photodiode type, in which the low-energy detector works both as absorption filter of low-energy fraction of radiation for the high-energy detector and detector of the low-energy radiation. The energy dependence deviation does not exceed ±5 % in the range from 30 to 120 keV. The developed detector and the instrument designed on its base allow contactless measurement of anode voltage of X-ray emitter from 40 to 140 kV with accuracy of ±3 %. The dose rate measurement range is 1-200 R/min.
issn 1027-5495
url https://nasplib.isofts.kiev.ua/handle/123456789/119702
citation_txt Detectors and instruments for detection of X-ray radiation on the basis of zinc selenide crystals / V.D. Ryzhikov, V.G. Volkov, O.D. Opolonin, S.V. Makhota, V.V. Maksymchuk // Functional Materials. — 2015. — Т. 22, № 4. — С. 552-558. — Бібліогр.: 18 назв. — англ.
work_keys_str_mv AT ryzhikovvd detectorsandinstrumentsfordetectionofxrayradiationonthebasisofzincselenidecrystals
AT volkovvg detectorsandinstrumentsfordetectionofxrayradiationonthebasisofzincselenidecrystals
AT opoloninod detectorsandinstrumentsfordetectionofxrayradiationonthebasisofzincselenidecrystals
AT makhotasv detectorsandinstrumentsfordetectionofxrayradiationonthebasisofzincselenidecrystals
AT maksymchukvv detectorsandinstrumentsfordetectionofxrayradiationonthebasisofzincselenidecrystals
first_indexed 2025-12-07T20:06:37Z
last_indexed 2025-12-07T20:06:37Z
_version_ 1850881340932620288