Effective thickness of the planar detector in measurements of electrons energy loss

An experimental method of determining the active region thickness of Si planar detector was used. The method based on the dependence the depletion layer thickness from voltage applied to the detector (U = 0...60 V ). The electron energy loss spectra emitted by ⁹⁰Sr -⁹⁰Y in silicon planar detector we...

Full description

Saved in:
Bibliographic Details
Published in:Вопросы атомной науки и техники
Date:2017
Main Authors: Deiev, O.S., Kiprich, S.K., Vasilyev, G.P., Yalovenko, V.I., Ovchinnik, V.D., Shulika, M.Y.
Format: Article
Language:English
Published: Національний науковий центр «Харківський фізико-технічний інститут» НАН України 2017
Subjects:
Online Access:https://nasplib.isofts.kiev.ua/handle/123456789/136099
Tags: Add Tag
No Tags, Be the first to tag this record!
Journal Title:Digital Library of Periodicals of National Academy of Sciences of Ukraine
Cite this:Effective thickness of the planar detector in measurements of electrons energy loss / O.S. Deiev, S.K. Kiprich, G.P. Vasilyev, V.I. Yalovenko, V.D. Ovchinnik, M.Y. Shulika // Физика низких температур. — 2017. — № 3. — С. 45-49. — Бібліогр.: 14 назв. — англ.

Institution

Digital Library of Periodicals of National Academy of Sciences of Ukraine
id nasplib_isofts_kiev_ua-123456789-136099
record_format dspace
spelling Deiev, O.S.
Kiprich, S.K.
Vasilyev, G.P.
Yalovenko, V.I.
Ovchinnik, V.D.
Shulika, M.Y.
2018-06-15T19:26:10Z
2018-06-15T19:26:10Z
2017
Effective thickness of the planar detector in measurements of electrons energy loss / O.S. Deiev, S.K. Kiprich, G.P. Vasilyev, V.I. Yalovenko, V.D. Ovchinnik, M.Y. Shulika // Физика низких температур. — 2017. — № 3. — С. 45-49. — Бібліогр.: 14 назв. — англ.
1562-6016
PACS: 07.85.Fv, 61.80.Cb
https://nasplib.isofts.kiev.ua/handle/123456789/136099
An experimental method of determining the active region thickness of Si planar detector was used. The method based on the dependence the depletion layer thickness from voltage applied to the detector (U = 0...60 V ). The electron energy loss spectra emitted by ⁹⁰Sr -⁹⁰Y in silicon planar detector were measured. The relative values of most probable energy loss of electrons were de ned for different thickness of detector. The planar detector was considered as a parallel-plate capacitor. The static (capacity) and dynamic (the detection efficiency and the energy deposit) characteristics had a root dependence on voltage.
Використовувався експериментальний метод визначення товщини активної області Si-планарного детектора. Метод заснований на залежності товщини збідненого шару від напруги, прикладеної до детектора (0...60 В). Вимірювалися спектри втрат енергії електронів, що випускаються ⁹⁰Sr -⁹⁰Y в кремнієвому планарному детекторі. Були визначені відносні значення найбільш ймовірних втрат енергії електронів для різної товщини детектора. Плоский детектор розглядався як конденсатор з паралельними пластинами. Статична (ємність) і динамічна (ефективність реєстрації та енергетичні втрати) характеристики мають кореневу залежністю від напруги.
Использовался экспериментальный метод определения толщины активной области Si-планарного детектора. Метод основан на зависимости толщины обедненного слоя от напряжения, приложенного к детектору (0...60 В). Измерялись спектры потерь энергии электронов, испускаемые ⁹⁰Sr -⁹⁰Y в кремниевом планарном детекторе. Были определены относительные значения наиболее вероятных потерь энергии электронов для различной толщины детектора. Плоский детектор рассматривался как конденсатор с параллельными пластинами. Статическая (емкость) и динамическая (эффективность регистрации и энергетические потери) характеристики имеют корневую зависимостью от напряжения.
en
Національний науковий центр «Харківський фізико-технічний інститут» НАН України
Вопросы атомной науки и техники
Ядерно-физические методы и обработка данных
Effective thickness of the planar detector in measurements of electrons energy loss
Ефективна товщина планарного детектора при вимірах втрати енергії електронів
Эфективная толщина планарного детектора при измерениях потери энергии электронов
Article
published earlier
institution Digital Library of Periodicals of National Academy of Sciences of Ukraine
collection DSpace DC
title Effective thickness of the planar detector in measurements of electrons energy loss
spellingShingle Effective thickness of the planar detector in measurements of electrons energy loss
Deiev, O.S.
Kiprich, S.K.
Vasilyev, G.P.
Yalovenko, V.I.
Ovchinnik, V.D.
Shulika, M.Y.
Ядерно-физические методы и обработка данных
title_short Effective thickness of the planar detector in measurements of electrons energy loss
title_full Effective thickness of the planar detector in measurements of electrons energy loss
title_fullStr Effective thickness of the planar detector in measurements of electrons energy loss
title_full_unstemmed Effective thickness of the planar detector in measurements of electrons energy loss
title_sort effective thickness of the planar detector in measurements of electrons energy loss
author Deiev, O.S.
Kiprich, S.K.
Vasilyev, G.P.
Yalovenko, V.I.
Ovchinnik, V.D.
Shulika, M.Y.
author_facet Deiev, O.S.
Kiprich, S.K.
Vasilyev, G.P.
Yalovenko, V.I.
Ovchinnik, V.D.
Shulika, M.Y.
topic Ядерно-физические методы и обработка данных
topic_facet Ядерно-физические методы и обработка данных
publishDate 2017
language English
container_title Вопросы атомной науки и техники
publisher Національний науковий центр «Харківський фізико-технічний інститут» НАН України
format Article
title_alt Ефективна товщина планарного детектора при вимірах втрати енергії електронів
Эфективная толщина планарного детектора при измерениях потери энергии электронов
description An experimental method of determining the active region thickness of Si planar detector was used. The method based on the dependence the depletion layer thickness from voltage applied to the detector (U = 0...60 V ). The electron energy loss spectra emitted by ⁹⁰Sr -⁹⁰Y in silicon planar detector were measured. The relative values of most probable energy loss of electrons were de ned for different thickness of detector. The planar detector was considered as a parallel-plate capacitor. The static (capacity) and dynamic (the detection efficiency and the energy deposit) characteristics had a root dependence on voltage. Використовувався експериментальний метод визначення товщини активної області Si-планарного детектора. Метод заснований на залежності товщини збідненого шару від напруги, прикладеної до детектора (0...60 В). Вимірювалися спектри втрат енергії електронів, що випускаються ⁹⁰Sr -⁹⁰Y в кремнієвому планарному детекторі. Були визначені відносні значення найбільш ймовірних втрат енергії електронів для різної товщини детектора. Плоский детектор розглядався як конденсатор з паралельними пластинами. Статична (ємність) і динамічна (ефективність реєстрації та енергетичні втрати) характеристики мають кореневу залежністю від напруги. Использовался экспериментальный метод определения толщины активной области Si-планарного детектора. Метод основан на зависимости толщины обедненного слоя от напряжения, приложенного к детектору (0...60 В). Измерялись спектры потерь энергии электронов, испускаемые ⁹⁰Sr -⁹⁰Y в кремниевом планарном детекторе. Были определены относительные значения наиболее вероятных потерь энергии электронов для различной толщины детектора. Плоский детектор рассматривался как конденсатор с параллельными пластинами. Статическая (емкость) и динамическая (эффективность регистрации и энергетические потери) характеристики имеют корневую зависимостью от напряжения.
issn 1562-6016
url https://nasplib.isofts.kiev.ua/handle/123456789/136099
citation_txt Effective thickness of the planar detector in measurements of electrons energy loss / O.S. Deiev, S.K. Kiprich, G.P. Vasilyev, V.I. Yalovenko, V.D. Ovchinnik, M.Y. Shulika // Физика низких температур. — 2017. — № 3. — С. 45-49. — Бібліогр.: 14 назв. — англ.
work_keys_str_mv AT deievos effectivethicknessoftheplanardetectorinmeasurementsofelectronsenergyloss
AT kiprichsk effectivethicknessoftheplanardetectorinmeasurementsofelectronsenergyloss
AT vasilyevgp effectivethicknessoftheplanardetectorinmeasurementsofelectronsenergyloss
AT yalovenkovi effectivethicknessoftheplanardetectorinmeasurementsofelectronsenergyloss
AT ovchinnikvd effectivethicknessoftheplanardetectorinmeasurementsofelectronsenergyloss
AT shulikamy effectivethicknessoftheplanardetectorinmeasurementsofelectronsenergyloss
AT deievos efektivnatovŝinaplanarnogodetektoraprivimírahvtratienergííelektronív
AT kiprichsk efektivnatovŝinaplanarnogodetektoraprivimírahvtratienergííelektronív
AT vasilyevgp efektivnatovŝinaplanarnogodetektoraprivimírahvtratienergííelektronív
AT yalovenkovi efektivnatovŝinaplanarnogodetektoraprivimírahvtratienergííelektronív
AT ovchinnikvd efektivnatovŝinaplanarnogodetektoraprivimírahvtratienergííelektronív
AT shulikamy efektivnatovŝinaplanarnogodetektoraprivimírahvtratienergííelektronív
AT deievos éfektivnaâtolŝinaplanarnogodetektorapriizmereniâhpoteriénergiiélektronov
AT kiprichsk éfektivnaâtolŝinaplanarnogodetektorapriizmereniâhpoteriénergiiélektronov
AT vasilyevgp éfektivnaâtolŝinaplanarnogodetektorapriizmereniâhpoteriénergiiélektronov
AT yalovenkovi éfektivnaâtolŝinaplanarnogodetektorapriizmereniâhpoteriénergiiélektronov
AT ovchinnikvd éfektivnaâtolŝinaplanarnogodetektorapriizmereniâhpoteriénergiiélektronov
AT shulikamy éfektivnaâtolŝinaplanarnogodetektorapriizmereniâhpoteriénergiiélektronov
first_indexed 2025-12-07T13:39:04Z
last_indexed 2025-12-07T13:39:04Z
_version_ 1850856958215585792