Radiation shield of therapeutic electron accelerator “CLINAC”

It is anticipated that a therapeutic electron accelerator “Clinac” (manufactured by Varian Co., USA) will be installed at NSC KIPT. It has been suggested that the available compartment having thick concrete walls might be used as a medical treatment room. In this connection, the analysis has been ma...

Full description

Saved in:
Bibliographic Details
Published in:Вопросы атомной науки и техники
Date:2017
Main Authors: Voronko, V.A., Dovbnya, A.N., Pugachev, G.D., Sotnikov, V.V., Uvarov, V.L., Shestakova, V.S.
Format: Article
Language:English
Published: Національний науковий центр «Харківський фізико-технічний інститут» НАН України 2017
Subjects:
Online Access:https://nasplib.isofts.kiev.ua/handle/123456789/136196
Tags: Add Tag
No Tags, Be the first to tag this record!
Journal Title:Digital Library of Periodicals of National Academy of Sciences of Ukraine
Cite this:Radiation shield of therapeutic electron accelerator “CLINAC” / V.A. Voronko, A.N. Dovbnya, G.D. Pugachev, V.V. Sotnikov, V.L. Uvarov, V.S. Shestakova // Вопросы атомной науки и техники. — 2017. — № 6. — С. 181-184. — Бібліогр.: 5 назв. — англ.

Institution

Digital Library of Periodicals of National Academy of Sciences of Ukraine
_version_ 1862619073522696192
author Voronko, V.A.
Dovbnya, A.N.
Pugachev, G.D.
Sotnikov, V.V.
Uvarov, V.L.
Shestakova, V.S.
author_facet Voronko, V.A.
Dovbnya, A.N.
Pugachev, G.D.
Sotnikov, V.V.
Uvarov, V.L.
Shestakova, V.S.
citation_txt Radiation shield of therapeutic electron accelerator “CLINAC” / V.A. Voronko, A.N. Dovbnya, G.D. Pugachev, V.V. Sotnikov, V.L. Uvarov, V.S. Shestakova // Вопросы атомной науки и техники. — 2017. — № 6. — С. 181-184. — Бібліогр.: 5 назв. — англ.
collection DSpace DC
container_title Вопросы атомной науки и техники
description It is anticipated that a therapeutic electron accelerator “Clinac” (manufactured by Varian Co., USA) will be installed at NSC KIPT. It has been suggested that the available compartment having thick concrete walls might be used as a medical treatment room. In this connection, the analysis has been made for correspondence of the com-partment protection characteristics to the accelerator safe operation requirements. The protection capability was estimated for a mode of the accelerator operation at electron energy of 18 MeV with a target made of tungsten. Equivalent dose rates behind the shield were calculated with due regard for the direct bremsstrahlung from the target, and also, for the scattered photon emission and neutron radiation. To reduce the scattered radiation level in the entrance area, the variant of shielding maze has been proposed. The estimated radiation dose of the accelerator staff does not exceed the values permissible by the radiation safety standards. According to the calculations made, the concentration of ozone produced in the treatment room will be also below the MPC level. У ННЦ ХФТІ передбачається встановити терапевтичний прискорювач електронів “Clinac” (фірма Varian, США). Як процедурний кабінет запропоновано використовувати наявне приміщення з товстими стінами із бетону. У зв'язку з цим проведено аналіз відповідності захисних характеристик приміщення вимогам безпечної експлуатації прискорювача. Оцінка захисту проводилася для режиму роботи прискорювача з енергією електронів 18 МеВ і мішенню із вольфраму. Розраховані значення потужності еквівалентної дози за захистом з урахуванням прямого гальмівного випромінювання від мішені, а також розсіяного фотонного випромінення і нейтронів. Для зниження рівня розсіяного випромінювання біля входу в приміщення запропонований варіант захисного лабіринту. Отримана оцінка дозового навантаження на обслуговуючий персонал прискорювача не перевищує значень, що допускаються нормами радіаційної безпеки. Проведений розрахунок напрацювання озону в процедурному кабінеті показав, що його концентрація також не перевищуватиме допустимий рівень. В ННЦ ХФТИ предполагается установить терапевтический ускоритель электронов “Clinac” (фирма Varian, США). В качестве процедурного кабинета предложено использовать имеющееся помещение с толстыми стенами из бетона. В связи с этим проведен анализ соответствия защитных характеристик помещения требованиям безопасной эксплуатации ускорителя. Оценка защиты проводилась для режима работы ускорителя с энергией электронов 18 МэВ и мишенью из вольфрама. Рассчитаны значения мощности эквивалентной дозы за защитой с учетом прямого тормозного излучения от мишени, а также рассеянного фотонного излучения и нейтронов. Для снижения уровня рассеянного излучения у входа в помещение предложен вариант защитного лабиринта. Полученная оценка дозовой нагрузки на обслуживающий персонал ускорителя не превышает значений, допускаемых нормами радиационной безопасности. Проведенный расчет наработки озона в процедурном кабинете показал, что его концентрация также не будет превышать допустимый уровень.
first_indexed 2025-12-07T13:16:27Z
format Article
fulltext
id nasplib_isofts_kiev_ua-123456789-136196
institution Digital Library of Periodicals of National Academy of Sciences of Ukraine
issn 1562-6016
language English
last_indexed 2025-12-07T13:16:27Z
publishDate 2017
publisher Національний науковий центр «Харківський фізико-технічний інститут» НАН України
record_format dspace
spelling Voronko, V.A.
Dovbnya, A.N.
Pugachev, G.D.
Sotnikov, V.V.
Uvarov, V.L.
Shestakova, V.S.
2018-06-16T07:04:12Z
2018-06-16T07:04:12Z
2017
Radiation shield of therapeutic electron accelerator “CLINAC” / V.A. Voronko, A.N. Dovbnya, G.D. Pugachev, V.V. Sotnikov, V.L. Uvarov, V.S. Shestakova // Вопросы атомной науки и техники. — 2017. — № 6. — С. 181-184. — Бібліогр.: 5 назв. — англ.
1562-6016
PACS: 87.56.bd; 28.41.Qb
https://nasplib.isofts.kiev.ua/handle/123456789/136196
It is anticipated that a therapeutic electron accelerator “Clinac” (manufactured by Varian Co., USA) will be installed at NSC KIPT. It has been suggested that the available compartment having thick concrete walls might be used as a medical treatment room. In this connection, the analysis has been made for correspondence of the com-partment protection characteristics to the accelerator safe operation requirements. The protection capability was estimated for a mode of the accelerator operation at electron energy of 18 MeV with a target made of tungsten. Equivalent dose rates behind the shield were calculated with due regard for the direct bremsstrahlung from the target, and also, for the scattered photon emission and neutron radiation. To reduce the scattered radiation level in the entrance area, the variant of shielding maze has been proposed. The estimated radiation dose of the accelerator staff does not exceed the values permissible by the radiation safety standards. According to the calculations made, the concentration of ozone produced in the treatment room will be also below the MPC level.
У ННЦ ХФТІ передбачається встановити терапевтичний прискорювач електронів “Clinac” (фірма Varian, США). Як процедурний кабінет запропоновано використовувати наявне приміщення з товстими стінами із бетону. У зв'язку з цим проведено аналіз відповідності захисних характеристик приміщення вимогам безпечної експлуатації прискорювача. Оцінка захисту проводилася для режиму роботи прискорювача з енергією електронів 18 МеВ і мішенню із вольфраму. Розраховані значення потужності еквівалентної дози за захистом з урахуванням прямого гальмівного випромінювання від мішені, а також розсіяного фотонного випромінення і нейтронів. Для зниження рівня розсіяного випромінювання біля входу в приміщення запропонований варіант захисного лабіринту. Отримана оцінка дозового навантаження на обслуговуючий персонал прискорювача не перевищує значень, що допускаються нормами радіаційної безпеки. Проведений розрахунок напрацювання озону в процедурному кабінеті показав, що його концентрація також не перевищуватиме допустимий рівень.
В ННЦ ХФТИ предполагается установить терапевтический ускоритель электронов “Clinac” (фирма Varian, США). В качестве процедурного кабинета предложено использовать имеющееся помещение с толстыми стенами из бетона. В связи с этим проведен анализ соответствия защитных характеристик помещения требованиям безопасной эксплуатации ускорителя. Оценка защиты проводилась для режима работы ускорителя с энергией электронов 18 МэВ и мишенью из вольфрама. Рассчитаны значения мощности эквивалентной дозы за защитой с учетом прямого тормозного излучения от мишени, а также рассеянного фотонного излучения и нейтронов. Для снижения уровня рассеянного излучения у входа в помещение предложен вариант защитного лабиринта. Полученная оценка дозовой нагрузки на обслуживающий персонал ускорителя не превышает значений, допускаемых нормами радиационной безопасности. Проведенный расчет наработки озона в процедурном кабинете показал, что его концентрация также не будет превышать допустимый уровень.
en
Національний науковий центр «Харківський фізико-технічний інститут» НАН України
Вопросы атомной науки и техники
Применение ускорителей в радиационных технологиях
Radiation shield of therapeutic electron accelerator “CLINAC”
Радіаційний захист терапевтичного прискорювача електронів “CLINAC”
Радиационная защита терапевтического ускорителя электронов «CLINAC»
Article
published earlier
spellingShingle Radiation shield of therapeutic electron accelerator “CLINAC”
Voronko, V.A.
Dovbnya, A.N.
Pugachev, G.D.
Sotnikov, V.V.
Uvarov, V.L.
Shestakova, V.S.
Применение ускорителей в радиационных технологиях
title Radiation shield of therapeutic electron accelerator “CLINAC”
title_alt Радіаційний захист терапевтичного прискорювача електронів “CLINAC”
Радиационная защита терапевтического ускорителя электронов «CLINAC»
title_full Radiation shield of therapeutic electron accelerator “CLINAC”
title_fullStr Radiation shield of therapeutic electron accelerator “CLINAC”
title_full_unstemmed Radiation shield of therapeutic electron accelerator “CLINAC”
title_short Radiation shield of therapeutic electron accelerator “CLINAC”
title_sort radiation shield of therapeutic electron accelerator “clinac”
topic Применение ускорителей в радиационных технологиях
topic_facet Применение ускорителей в радиационных технологиях
url https://nasplib.isofts.kiev.ua/handle/123456789/136196
work_keys_str_mv AT voronkova radiationshieldoftherapeuticelectronacceleratorclinac
AT dovbnyaan radiationshieldoftherapeuticelectronacceleratorclinac
AT pugachevgd radiationshieldoftherapeuticelectronacceleratorclinac
AT sotnikovvv radiationshieldoftherapeuticelectronacceleratorclinac
AT uvarovvl radiationshieldoftherapeuticelectronacceleratorclinac
AT shestakovavs radiationshieldoftherapeuticelectronacceleratorclinac
AT voronkova radíacíiniizahistterapevtičnogopriskorûvačaelektronívclinac
AT dovbnyaan radíacíiniizahistterapevtičnogopriskorûvačaelektronívclinac
AT pugachevgd radíacíiniizahistterapevtičnogopriskorûvačaelektronívclinac
AT sotnikovvv radíacíiniizahistterapevtičnogopriskorûvačaelektronívclinac
AT uvarovvl radíacíiniizahistterapevtičnogopriskorûvačaelektronívclinac
AT shestakovavs radíacíiniizahistterapevtičnogopriskorûvačaelektronívclinac
AT voronkova radiacionnaâzaŝitaterapevtičeskogouskoritelâélektronovclinac
AT dovbnyaan radiacionnaâzaŝitaterapevtičeskogouskoritelâélektronovclinac
AT pugachevgd radiacionnaâzaŝitaterapevtičeskogouskoritelâélektronovclinac
AT sotnikovvv radiacionnaâzaŝitaterapevtičeskogouskoritelâélektronovclinac
AT uvarovvl radiacionnaâzaŝitaterapevtičeskogouskoritelâélektronovclinac
AT shestakovavs radiacionnaâzaŝitaterapevtičeskogouskoritelâélektronovclinac