Potentiality of application of the conductometric L-arginine biosensors for the real sample analysis

Aim. To determine an influence of serum components on the L-arginine biosensor sensitivity and to formulate
 practical recommendations for its reliable analysis. Methods. The L-arginine biosensor comprised arginase and
 urease co-immobilized by cross-linking. Results. The biosensor s...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Вiopolymers and Cell
Datum:2012
Hauptverfasser: Saiapina, O.Y., Dzyadevych, S.V., Jaffrezic-Renault, N.
Format: Artikel
Sprache:Englisch
Veröffentlicht: Інститут молекулярної біології і генетики НАН України 2012
Schlagworte:
Online Zugang:https://nasplib.isofts.kiev.ua/handle/123456789/156756
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Назва журналу:Digital Library of Periodicals of National Academy of Sciences of Ukraine
Zitieren:Potentiality of application of the conductometric
 L-arginine biosensors for the real sample analysis / O.Y. Saiapina, S.V. Dzyadevych, N. Jaffrezic-Renault // Вiopolymers and Cell. — 2012. — Т. 28, № 6. — С. 441-448. — Бібліогр.: 27 назв. — англ.

Institution

Digital Library of Periodicals of National Academy of Sciences of Ukraine
_version_ 1862545375053742080
author Saiapina, O.Y.
Dzyadevych, S.V.
Jaffrezic-Renault, N.
author_facet Saiapina, O.Y.
Dzyadevych, S.V.
Jaffrezic-Renault, N.
citation_txt Potentiality of application of the conductometric
 L-arginine biosensors for the real sample analysis / O.Y. Saiapina, S.V. Dzyadevych, N. Jaffrezic-Renault // Вiopolymers and Cell. — 2012. — Т. 28, № 6. — С. 441-448. — Бібліогр.: 27 назв. — англ.
collection DSpace DC
container_title Вiopolymers and Cell
description Aim. To determine an influence of serum components on the L-arginine biosensor sensitivity and to formulate
 practical recommendations for its reliable analysis. Methods. The L-arginine biosensor comprised arginase and
 urease co-immobilized by cross-linking. Results. The biosensor specificity was investigated based on a series of
 representative studies (namely, through urea determination in the serum; inhibitory effect studies of mercury
 ions; high temperature treatment of sensors; studying the biosensor sensitivity to the serum treated by enzymes,
 and selectivity studies). It was found that the response of the biosensor to the serum injections was determined by
 high sensitivity of the L-arginine biosensor toward not only to L-arginine but also toward two other basic amino
 acids (L-lysine and L-histidine). Conclusions. A detailed procedure of optimization of the conductometric biosensor for L-arginine determination in blood serum has been proposed.
 Keywords: L-arginine, conductometric biosensors, serum, optimization procedure. Мета. Визначити вплив компонентів сироватки крові на чутливість біосенсора при виявленні L-аргініну та сформулювати практичні рекомендації для забезпечення її надійного аналізу. Методи. Біосенсор для визначення L-аргініну містить аргіназу і уреазу, коіммобілізовані методом поперечного зшивання. Результати. Специфічність біосенсора вивчали на основі низки показників – вмісту сечовини у сироватці; інгібувального впливу іонів ртуті; високотемпературної обробки біосенсорів; чутливості біосенсора до сироватки крові, обробленої ліофілізованими препаратами ферментів, та селективності біосенсора. Встановлено, що відгук біосенсора на внесення сироватки зумовлений високою чутливістю біосенсора ще до двох, крім L-аргініну, основних амінокислот (L-лізину та L-гістидину). Висновки. Запропоновано детальну процедуру оптимізації кондуктометричного біосенсора для визначення L-аргініну у сироватці крові.
 Ключові слова: L-аргінін, кондуктометричні біосенсори, сироватка крові, процедура оптимізації. Цель. Определить влияние компонентов сыворотки крови на чувствительность биосенсора для выявления L-аргинина и сформулировать практические рекомендации для обеспечения ее надежного анализа. Методы. Биосенсор для определения L-аргинина содержит аргиназу и уреазу, ко-иммобилизованные методом поперечной сшивки. Результаты. Специфичность биосенсора изучали на основе серии показателей – содержания мочевины в сыворотке; ингибирующего эффекта ионов ртути; высокотемпературной обработки биосенсоров; чувствительности биосенсора к сыворотке крови, обработанной лиофилизованными препаратами ферментов, и селективности биосенсора. Установлено, что отклик биосенсора на внесение сыворотки обусловлен высокой чувствительностью биосенсора еще к двум, кроме L-аргинина, основным аминокислотам (L-лизину и L-гистидину). Выводы. Предложена детальная процедура оптимизации кондуктометрического биосенсора для определения L-аргинина в сыворотке крови.
 Ключевые слова: L-аргинин, кондуктометрические биосенсоры, сыворотка крови, процедура оптимизации.
first_indexed 2025-11-25T07:20:28Z
format Article
fulltext
id nasplib_isofts_kiev_ua-123456789-156756
institution Digital Library of Periodicals of National Academy of Sciences of Ukraine
issn 0233-7657
language English
last_indexed 2025-11-25T07:20:28Z
publishDate 2012
publisher Інститут молекулярної біології і генетики НАН України
record_format dspace
spelling Saiapina, O.Y.
Dzyadevych, S.V.
Jaffrezic-Renault, N.
2019-06-18T20:12:57Z
2019-06-18T20:12:57Z
2012
Potentiality of application of the conductometric
 L-arginine biosensors for the real sample analysis / O.Y. Saiapina, S.V. Dzyadevych, N. Jaffrezic-Renault // Вiopolymers and Cell. — 2012. — Т. 28, № 6. — С. 441-448. — Бібліогр.: 27 назв. — англ.
0233-7657
DOI: http://dx.doi.org/10.7124/bc.000134
https://nasplib.isofts.kiev.ua/handle/123456789/156756
543.555 + 551.508.91 + 577.112.385
Aim. To determine an influence of serum components on the L-arginine biosensor sensitivity and to formulate
 practical recommendations for its reliable analysis. Methods. The L-arginine biosensor comprised arginase and
 urease co-immobilized by cross-linking. Results. The biosensor specificity was investigated based on a series of
 representative studies (namely, through urea determination in the serum; inhibitory effect studies of mercury
 ions; high temperature treatment of sensors; studying the biosensor sensitivity to the serum treated by enzymes,
 and selectivity studies). It was found that the response of the biosensor to the serum injections was determined by
 high sensitivity of the L-arginine biosensor toward not only to L-arginine but also toward two other basic amino
 acids (L-lysine and L-histidine). Conclusions. A detailed procedure of optimization of the conductometric biosensor for L-arginine determination in blood serum has been proposed.
 Keywords: L-arginine, conductometric biosensors, serum, optimization procedure.
Мета. Визначити вплив компонентів сироватки крові на чутливість біосенсора при виявленні L-аргініну та сформулювати практичні рекомендації для забезпечення її надійного аналізу. Методи. Біосенсор для визначення L-аргініну містить аргіназу і уреазу, коіммобілізовані методом поперечного зшивання. Результати. Специфічність біосенсора вивчали на основі низки показників – вмісту сечовини у сироватці; інгібувального впливу іонів ртуті; високотемпературної обробки біосенсорів; чутливості біосенсора до сироватки крові, обробленої ліофілізованими препаратами ферментів, та селективності біосенсора. Встановлено, що відгук біосенсора на внесення сироватки зумовлений високою чутливістю біосенсора ще до двох, крім L-аргініну, основних амінокислот (L-лізину та L-гістидину). Висновки. Запропоновано детальну процедуру оптимізації кондуктометричного біосенсора для визначення L-аргініну у сироватці крові.
 Ключові слова: L-аргінін, кондуктометричні біосенсори, сироватка крові, процедура оптимізації.
Цель. Определить влияние компонентов сыворотки крови на чувствительность биосенсора для выявления L-аргинина и сформулировать практические рекомендации для обеспечения ее надежного анализа. Методы. Биосенсор для определения L-аргинина содержит аргиназу и уреазу, ко-иммобилизованные методом поперечной сшивки. Результаты. Специфичность биосенсора изучали на основе серии показателей – содержания мочевины в сыворотке; ингибирующего эффекта ионов ртути; высокотемпературной обработки биосенсоров; чувствительности биосенсора к сыворотке крови, обработанной лиофилизованными препаратами ферментов, и селективности биосенсора. Установлено, что отклик биосенсора на внесение сыворотки обусловлен высокой чувствительностью биосенсора еще к двум, кроме L-аргинина, основным аминокислотам (L-лизину и L-гистидину). Выводы. Предложена детальная процедура оптимизации кондуктометрического биосенсора для определения L-аргинина в сыворотке крови.
 Ключевые слова: L-аргинин, кондуктометрические биосенсоры, сыворотка крови, процедура оптимизации.
The authors would like to thank
 National Academy of Sciences of Ukraine (complex scientific-technical program «Sensor systems for medical-ecological and industrial purposes»), the European
 Commission for their funding of the Project IRSESNANODEV and Rhone-Alpes Region for MIRA project.
en
Інститут молекулярної біології і генетики НАН України
Вiopolymers and Cell
Molecular and Cell Biotechnologies
Potentiality of application of the conductometric L-arginine biosensors for the real sample analysis
Потенційна можливість застосування кондуктометричних біосенсорів для визначення L-аргініну при аналізі реальних зразків
Потенциальная возможность применения кондуктометрических биосенсоров для определения L-аргинина при анализе реальных образцов
Article
published earlier
spellingShingle Potentiality of application of the conductometric L-arginine biosensors for the real sample analysis
Saiapina, O.Y.
Dzyadevych, S.V.
Jaffrezic-Renault, N.
Molecular and Cell Biotechnologies
title Potentiality of application of the conductometric L-arginine biosensors for the real sample analysis
title_alt Потенційна можливість застосування кондуктометричних біосенсорів для визначення L-аргініну при аналізі реальних зразків
Потенциальная возможность применения кондуктометрических биосенсоров для определения L-аргинина при анализе реальных образцов
title_full Potentiality of application of the conductometric L-arginine biosensors for the real sample analysis
title_fullStr Potentiality of application of the conductometric L-arginine biosensors for the real sample analysis
title_full_unstemmed Potentiality of application of the conductometric L-arginine biosensors for the real sample analysis
title_short Potentiality of application of the conductometric L-arginine biosensors for the real sample analysis
title_sort potentiality of application of the conductometric l-arginine biosensors for the real sample analysis
topic Molecular and Cell Biotechnologies
topic_facet Molecular and Cell Biotechnologies
url https://nasplib.isofts.kiev.ua/handle/123456789/156756
work_keys_str_mv AT saiapinaoy potentialityofapplicationoftheconductometriclargininebiosensorsfortherealsampleanalysis
AT dzyadevychsv potentialityofapplicationoftheconductometriclargininebiosensorsfortherealsampleanalysis
AT jaffrezicrenaultn potentialityofapplicationoftheconductometriclargininebiosensorsfortherealsampleanalysis
AT saiapinaoy potencíinamožlivístʹzastosuvannâkonduktometričnihbíosensorívdlâviznačennâlargínínuprianalízírealʹnihzrazkív
AT dzyadevychsv potencíinamožlivístʹzastosuvannâkonduktometričnihbíosensorívdlâviznačennâlargínínuprianalízírealʹnihzrazkív
AT jaffrezicrenaultn potencíinamožlivístʹzastosuvannâkonduktometričnihbíosensorívdlâviznačennâlargínínuprianalízírealʹnihzrazkív
AT saiapinaoy potencialʹnaâvozmožnostʹprimeneniâkonduktometričeskihbiosensorovdlâopredeleniâlargininaprianalizerealʹnyhobrazcov
AT dzyadevychsv potencialʹnaâvozmožnostʹprimeneniâkonduktometričeskihbiosensorovdlâopredeleniâlargininaprianalizerealʹnyhobrazcov
AT jaffrezicrenaultn potencialʹnaâvozmožnostʹprimeneniâkonduktometričeskihbiosensorovdlâopredeleniâlargininaprianalizerealʹnyhobrazcov