Empirical formula for the dependence of HPGe-detector efficiency on energy and distance for shielded gamma-ray sources

The accuracy of gamma-spectrometric measurements in the isotopic analysis of shielded nuclear materials depends on the accuracy of the detector calibration in terms of energy efficiency, which should take into account corrections related to the measurement geometry and the absorption of γ-radiation...

Full description

Saved in:
Bibliographic Details
Published in:Problems of Atomic Science and Technology
Date:2023
Main Authors: Pylypchynets, I.V., Lengyel, O.I., Parlag, O.O., Oleynikov, E.V.
Format: Article
Language:English
Published: Національний науковий центр «Харківський фізико-технічний інститут» НАН України 2023
Subjects:
Online Access:https://nasplib.isofts.kiev.ua/handle/123456789/196137
Tags: Add Tag
No Tags, Be the first to tag this record!
Journal Title:Digital Library of Periodicals of National Academy of Sciences of Ukraine
Cite this:Empirical formula for the dependence of HPGe-detector efficiency on energy and distance for shielded gamma-ray sources / I.V. Pylypchynets, O.I. Lengyel, O.O. Parlag, E.V. Oleynikov // Problems of Atomic Science and Technology. — 2023. — № 3. — С. 47-51. — Бібліогр.: 22 назв. — англ.

Institution

Digital Library of Periodicals of National Academy of Sciences of Ukraine
_version_ 1862733570180644864
author Pylypchynets, I.V.
Lengyel, O.I.
Parlag, O.O.
Oleynikov, E.V.
author_facet Pylypchynets, I.V.
Lengyel, O.I.
Parlag, O.O.
Oleynikov, E.V.
citation_txt Empirical formula for the dependence of HPGe-detector efficiency on energy and distance for shielded gamma-ray sources / I.V. Pylypchynets, O.I. Lengyel, O.O. Parlag, E.V. Oleynikov // Problems of Atomic Science and Technology. — 2023. — № 3. — С. 47-51. — Бібліогр.: 22 назв. — англ.
collection DSpace DC
container_title Problems of Atomic Science and Technology
description The accuracy of gamma-spectrometric measurements in the isotopic analysis of shielded nuclear materials depends on the accuracy of the detector calibration in terms of energy efficiency, which should take into account corrections related to the measurement geometry and the absorption of γ-radiation by the shield material. The results of experimental studies of the energy efficiency of the HPGe-detector measured at fixed distances between the calibration source and the detector (50 and 100 mm) in the presence and absence of an absorbing shield made of stainless steel 12X18H10T (thickness ‒ 9.6 mm) are presented. On the basis of experimental data, an empirical description of the efficiency dependence for fixed distances between the gamma radiation source and the detector in the presence of a stainless steel absorbing shield was obtained. Точність гамма-спектрометричних вимірів при ізотопному аналізі екранованих ядерних матеріалів залежить від точності калібрування детектора за енергетичною ефективністю, що повинна враховувати поправки, пов’язані з геометрією вимірювань, та поглинанням γ-випромінювання матеріалом екрана. Представленo результати експериментальних досліджень енергетичної ефективності HPGe-детектора, яка виміряна при фіксованих відстанях калібрувальне джерело – детектор (50 і 100 мм) при наявності та відсутності поглинаючого екрану із нержавіючої сталі марки 12Х18Н10Т (товщина – 9,6 мм). На основі експериментальних даних отримано емпіричне описання залежності ефективності для фіксованих відстаней між джерелом гамма-випромінювання і детектором при наявності поглинаючого екрана із нержавіючої сталі.
first_indexed 2025-12-07T19:38:01Z
format Article
fulltext
id nasplib_isofts_kiev_ua-123456789-196137
institution Digital Library of Periodicals of National Academy of Sciences of Ukraine
issn 1562-6016
language English
last_indexed 2025-12-07T19:38:01Z
publishDate 2023
publisher Національний науковий центр «Харківський фізико-технічний інститут» НАН України
record_format dspace
spelling Pylypchynets, I.V.
Lengyel, O.I.
Parlag, O.O.
Oleynikov, E.V.
2023-12-10T16:49:35Z
2023-12-10T16:49:35Z
2023
Empirical formula for the dependence of HPGe-detector efficiency on energy and distance for shielded gamma-ray sources / I.V. Pylypchynets, O.I. Lengyel, O.O. Parlag, E.V. Oleynikov // Problems of Atomic Science and Technology. — 2023. — № 3. — С. 47-51. — Бібліогр.: 22 назв. — англ.
1562-6016
PACS: 24.75.+1, 25.85.-w, 25.85.Ec, 25.85. Ca
DOI: https://doi.org/10.46813/2023-145-047
https://nasplib.isofts.kiev.ua/handle/123456789/196137
The accuracy of gamma-spectrometric measurements in the isotopic analysis of shielded nuclear materials depends on the accuracy of the detector calibration in terms of energy efficiency, which should take into account corrections related to the measurement geometry and the absorption of γ-radiation by the shield material. The results of experimental studies of the energy efficiency of the HPGe-detector measured at fixed distances between the calibration source and the detector (50 and 100 mm) in the presence and absence of an absorbing shield made of stainless steel 12X18H10T (thickness ‒ 9.6 mm) are presented. On the basis of experimental data, an empirical description of the efficiency dependence for fixed distances between the gamma radiation source and the detector in the presence of a stainless steel absorbing shield was obtained.
Точність гамма-спектрометричних вимірів при ізотопному аналізі екранованих ядерних матеріалів залежить від точності калібрування детектора за енергетичною ефективністю, що повинна враховувати поправки, пов’язані з геометрією вимірювань, та поглинанням γ-випромінювання матеріалом екрана. Представленo результати експериментальних досліджень енергетичної ефективності HPGe-детектора, яка виміряна при фіксованих відстанях калібрувальне джерело – детектор (50 і 100 мм) при наявності та відсутності поглинаючого екрану із нержавіючої сталі марки 12Х18Н10Т (товщина – 9,6 мм). На основі експериментальних даних отримано емпіричне описання залежності ефективності для фіксованих відстаней між джерелом гамма-випромінювання і детектором при наявності поглинаючого екрана із нержавіючої сталі.
The authors express their gratitude to the microtron group of I.M. Kushtan for the technical support of the experimental studies.
en
Національний науковий центр «Харківський фізико-технічний інститут» НАН України
Problems of Atomic Science and Technology
Detectors and nuclear radiation detection
Empirical formula for the dependence of HPGe-detector efficiency on energy and distance for shielded gamma-ray sources
Емпірична формула залежності ефективності HPGe-детектора від енергії та відстані для екранованих джерел гамма-випромінювання
Article
published earlier
spellingShingle Empirical formula for the dependence of HPGe-detector efficiency on energy and distance for shielded gamma-ray sources
Pylypchynets, I.V.
Lengyel, O.I.
Parlag, O.O.
Oleynikov, E.V.
Detectors and nuclear radiation detection
title Empirical formula for the dependence of HPGe-detector efficiency on energy and distance for shielded gamma-ray sources
title_alt Емпірична формула залежності ефективності HPGe-детектора від енергії та відстані для екранованих джерел гамма-випромінювання
title_full Empirical formula for the dependence of HPGe-detector efficiency on energy and distance for shielded gamma-ray sources
title_fullStr Empirical formula for the dependence of HPGe-detector efficiency on energy and distance for shielded gamma-ray sources
title_full_unstemmed Empirical formula for the dependence of HPGe-detector efficiency on energy and distance for shielded gamma-ray sources
title_short Empirical formula for the dependence of HPGe-detector efficiency on energy and distance for shielded gamma-ray sources
title_sort empirical formula for the dependence of hpge-detector efficiency on energy and distance for shielded gamma-ray sources
topic Detectors and nuclear radiation detection
topic_facet Detectors and nuclear radiation detection
url https://nasplib.isofts.kiev.ua/handle/123456789/196137
work_keys_str_mv AT pylypchynetsiv empiricalformulaforthedependenceofhpgedetectorefficiencyonenergyanddistanceforshieldedgammaraysources
AT lengyeloi empiricalformulaforthedependenceofhpgedetectorefficiencyonenergyanddistanceforshieldedgammaraysources
AT parlagoo empiricalformulaforthedependenceofhpgedetectorefficiencyonenergyanddistanceforshieldedgammaraysources
AT oleynikovev empiricalformulaforthedependenceofhpgedetectorefficiencyonenergyanddistanceforshieldedgammaraysources
AT pylypchynetsiv empíričnaformulazaležnostíefektivnostíhpgedetektoravídenergíítavídstanídlâekranovanihdžerelgammavipromínûvannâ
AT lengyeloi empíričnaformulazaležnostíefektivnostíhpgedetektoravídenergíítavídstanídlâekranovanihdžerelgammavipromínûvannâ
AT parlagoo empíričnaformulazaležnostíefektivnostíhpgedetektoravídenergíítavídstanídlâekranovanihdžerelgammavipromínûvannâ
AT oleynikovev empíričnaformulazaležnostíefektivnostíhpgedetektoravídenergíítavídstanídlâekranovanihdžerelgammavipromínûvannâ