Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields

Results of a study of dosimetry characteristics of a prototype of the detector based on a polycrystalline diamond film (CVD-diamond) produced in NSC KIPT are summarized. The techniques of the detector calibration against the electron and X-ray radiation dose rate are developed. The conditions of c...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Вопросы атомной науки и техники
Datum:2014
Hauptverfasser: Kutny, V.E., Nikiforov, V.I., Opalev, O.A., Rogov, Yu.V., Rybka, A.V., Shevchenko, V.A., Shlyakhov, I.N., Strelnitsky, V.E., Tenishev, A.Eh., Uvarov, V.L.
Format: Artikel
Sprache:Englisch
Veröffentlicht: Національний науковий центр «Харківський фізико-технічний інститут» НАН України 2014
Schlagworte:
Online Zugang:https://nasplib.isofts.kiev.ua/handle/123456789/80276
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Назва журналу:Digital Library of Periodicals of National Academy of Sciences of Ukraine
Zitieren:Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields / V.E. Kutny, V.I. Nikiforov, O.A. Opalev, Yu.V. Rogov, A.V. Rybka, V.A. Shevchenko, I.N. Shlyakhov, V.E. Strelnitsky, A.Eh. Tenishev, V.L. Uvarov // Вопросы атомной науки и техники. — 2014. — № 3. — С. 162-165. — Бібліогр.: 12 назв. — англ.

Institution

Digital Library of Periodicals of National Academy of Sciences of Ukraine
_version_ 1859821594212302848
author Kutny, V.E.
Nikiforov, V.I.
Opalev, O.A.
Rogov, Yu.V.
Rybka, A.V.
Shevchenko, V.A.
Shlyakhov, I.N.
Strelnitsky, V.E.
Tenishev, A.Eh.
Uvarov, V.L.
author_facet Kutny, V.E.
Nikiforov, V.I.
Opalev, O.A.
Rogov, Yu.V.
Rybka, A.V.
Shevchenko, V.A.
Shlyakhov, I.N.
Strelnitsky, V.E.
Tenishev, A.Eh.
Uvarov, V.L.
citation_txt Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields / V.E. Kutny, V.I. Nikiforov, O.A. Opalev, Yu.V. Rogov, A.V. Rybka, V.A. Shevchenko, I.N. Shlyakhov, V.E. Strelnitsky, A.Eh. Tenishev, V.L. Uvarov // Вопросы атомной науки и техники. — 2014. — № 3. — С. 162-165. — Бібліогр.: 12 назв. — англ.
collection DSpace DC
container_title Вопросы атомной науки и техники
description Results of a study of dosimetry characteristics of a prototype of the detector based on a polycrystalline diamond film (CVD-diamond) produced in NSC KIPT are summarized. The techniques of the detector calibration against the electron and X-ray radiation dose rate are developed. The conditions of calibration were studied by means of computer simulation. For determination of detector sensitivity to electron radiation, it was placed inside of a standard polystyrene phantom. An additional filter of electrons was used at measurement in the X-ray field. Every time detector irradiation was carried out together with the Harwell Red Perpex 4034 dosimeters, which registrations provided the calibration data. The measurements, executed at the LU-10, EPOS and LU-40 linacs of NSC KIPT, have demonstrated, that the values of detector sensitivity against each type of radiation are close and conform to ones obtained earlier for the low-intensity radiation fields. Considering significant radiation durability of the CVD-diamond, it enables its application in technological dosimetry. Описаны результаты исследования дозиметрических характеристик изготовленного в ННЦ ХФТИ опытного образца детектора на основе поликристаллической алмазной пленки (CVD-алмаз). Разработаны методики калибровки детекторов по мощности поглощенной дозы электронного и тормозного излучений. Условия калибровки исследованы методом компьютерного моделирования. Для определения чувствительности детектора к электронному излучению его помещали внутри фантома из полистирола. При измерениях в поле тормозного излучения использовался дополнительный фильтр электронов. В каждом случае облучение детектора выполнялось совместно с дозиметрами Harwell Red Perspex 4034, по показаниям которых производилась калибровка. Проведенные на ускорителях ЛУ-10, ЭПОС и ЛУ-40 ННЦ ХФТИ измерения показали, что значения чувствительности детектора для каждого вида излучения близки и согласуются с полученными ранее в условиях полей излучения низкой интенсивности. Учитывая высокую радиационную стойкость CVD-алмаза, это обеспечивает возможность его применения в технологической дозиметрии. Описанo результати дослідження дозиметричних характеристик виготовленого в ННЦ ХФТІ дослідного зразка детектора на основі полікристалічної алмазної плівки (CVD-алмаз). Розроблено методики калібрування детекторів за потужністю поглинутої дози електронного та гальмівного випромінювань. Методом комп'ютерного моделювання досліджені умови калібрувань. Для визначення чутливості детектора до електронного випромінювання його розміщували усередині стандартного фантома з полістиролу. При вимірюваннях у полі гальмівного випромінювання використовувався додатковий фільтр електронів. У кожному випадку опромінювання детектора виконувалося спільно з дозиметрами Harwell Red Perpex 4034, за показаннями яких проводилося калібрування. Проведені на прискорювачах ЛУ-10, ЕПОС та ЛУ-40 ННЦ ХФТІ вимірювання показали, що значення чутливості детектора для кожного виду випромінювання є близькі і узгоджуються з такими, що одержані раніше в умовах полів випромінювання низької інтенсивності. Зважаючи на значну радіаційну стійкість CVD-алмазу, це забезпечує можливість його застосування в технологічній дозиметрії.
first_indexed 2025-12-07T15:25:52Z
format Article
fulltext ISSN 1562-6016. ВАНТ. 2014. №3(91) 162 CALIBRATION OF CVD-DIAMOND BASED DOSIMETER IN HIGH-POWER ELECTRON AND X-RAY RADIATION FIELDS V.E. Kutny, V.I. Nikiforov, O.A. Opalev, Yu.V. Rogov, A.V. Rybka, V.A. Shevchenko, I.N. Shlyakhov, V.E. Strelnitsky, A.Eh. Tenishev, V.L. Uvarov National Science Center “Kharkov Institute of Physics and Technology”, Kharkov, Ukraine E-mail: uvarov@kipt.kharkov.ua Results of a study of dosimetry characteristics of a prototype of the detector based on a polycrystalline diamond film (CVD-diamond) produced in NSC KIPT are summarized. The techniques of the detector calibration against the electron and X-ray radiation dose rate are developed. The conditions of calibration were studied by means of com- puter simulation. For determination of detector sensitivity to electron radiation, it was placed inside of a standard polystyrene phantom. An additional filter of electrons was used at measurement in the X-ray field. Every time detec- tor irradiation was carried out together with the Harwell Red Perpex 4034 dosimeters, which registrations provided the calibration data. The measurements, executed at the LU-10, EPOS and LU-40 linacs of NSC KIPT, have demon- strated, that the values of detector sensitivity against each type of radiation are close and conform to ones obtained earlier for the low-intensity radiation fields. Considering significant radiation durability of the CVD-diamond, it enables its application in technological dosimetry. PACS: 07.57.Kp, 29.40.-n, 85.25.Pb, 81.05.ug INTRODUCTION Natural diamond has a unique combination of elec- trophysical properties: wide gap (Еg=5.5 eV), high elec- trical strength (> 107 V/cm) and specific resistance (> 1011 Ohm·cm), and also, large energy of atom dis- placement from lattice (43 eV). These features allow to use diamond in radiation and nuclear physics, in par- ticular in detectors of γ- and β-radiation (see, for exam- ple, [1]). An additional possibility of diamond applica- tion in radiometry has appeared with the development of synthesis of the polycrystalline diamond films (PDF) and wafers by a method of chemical vapor deposition (CVD). This method has allowed to realize a large-scale manufacturing of detectors with the specific parameters for application in nuclear instrumentation [2], high en- ergy physics [3], medicine [4], etc. The data obtained recently on radiation hardness of CVD-diamond detectors [5] open also perspectives for their application in dosimetry of electron and X-Ray (bremsstrahlung) radiation at industrial accelerators. A specificity of such installations lays in the large average (> 10 kW) and pulsed (up to 10 МW) power of an elec- tron beam, determining the range of radiation and ther- mal loads on the detector. In this paper, the results of the measurement of analog sensitivity of the CVD- diamond detectors, developed in NSC KIPT, in the fields of the high-power electron and bremsstrahlung radiation are discussed. 1. METHODS AND MATERIALS 1.1. The PDF films for detectors were synthesised by the CVD method from a gas phase by activation of a carbon-hydrogen mixture with the glow dicharge, stabilized by a magnetic field. The samples obtained had the area up to 1 cm2 and thickness up to 350 microns. The quality of PDFs was checked by a method of infra-red spectroscopy. The parametrization of films was carried out by a measurement of their current-to-voltage characteristics. These measurements were made after each operation of annealing, etching, and exposure to radiation. With a specific electrical resistance ≥ 1012 Ohm·cm, samples meet the requirements, which the material for detector should comply with. In the process of detector manufacturing, its growth and substrate surfaces were polishied. The metal contacts (Al, Ti) of various geometry (strip, planar, coplanar) were then applied by the plasmochemical and photolytographic techniques. 1.2. The sample of the diamond detector (DD) (Fig. 1), used in experiments, was grown on a n-type silicon substrate of 1.5 mm thick. The thickness of PDF was 180 microns. On its growth surface, the two Al contacts were evaporated and used as a single common contact. The substrate was an another contact. The total area of the contact (the detector surface) was 10 mm2. The dark resistance of the detector before the exposure was 1.2·1011 Ohm. Fig. 1. Diamond detector 1.3. During the calibration measurements, the irradi- ation of the detector was monitored using the Red Per- spex 4034 dosimeters [6]. Commonly, they are used as the routine dosimetry systems in radiation technology with the use of electron and gamma radiation in the dose range 5…50 kGy. In our case, they were used for de- termination of a dose absorbed in DD. 1.4. Considering the high pulsed dose rate (~106 Gy/s), a study of the DD sensitivity under electron irradiation was carried out in an analog mode (Fig. 2). Detector Integrator Voltmeter Bias 100V Fig. 2. Coupling of DD at exposure to electron beam In the measurements with DD, a bias voltage of 100 V from an external source was applied to the detec- mailto:uvarov@kipt.kharkov.ua ISSN 1562-6016. ВАНТ. 2014. №3(91) 163 tor. The integral action time was 1 s. The integrator out- put voltage was measured with a digital voltmeter В7- 40. After a beam cut off, its value did not exceed 0.3 mV. 1.5. Determination of the DD sensitivity to X-Ray was carried out in a pulse mode (Fig. 3). The voltage drop at a load resistance R1 was measured with a digital oscilloscope TDS1012. Bias 100V DD R1 100kΩ TDS1012 Fig. 3. Coupling of DD at exposure to X-ray 1.6. The analysis of an irradiation field, as well as of its interaction with the elements of measuring devices was carried out using a computer simulation technique based on the program system PENELOPE-2008 [7]. 2. ELECTRON RADIATION 2.1. Calibration in a field of the electron radiation was carried out at an accelerator LU-10 in NSC KIPT [8] (Fig. 4). The DD was placed inside a standard poly- styrene RISO phantom [9], positioned at a distance of 110 cm from the outlet window of the accelerator. The phantom corresponds a parallelepiped 1 from foam pol- ystyrene measuring 29×29×10 см, in which centre, an absorber 2 in form of a disk from polystyrene 13.8 cm in diameter and 1.8 cm thick is placed. In the median plane of the disk, there is a 3 mm wide cavity 3, in which, the DD was positioned. In the same cavity, the routine dosimeters Red Perspex (RP) were placed. Their calibration was preliminarily conducted using the refer- ence polystyrene calorimetric dosimeters RISO, similar in structure to the phantoms [9]. e- LU-10 1 2 3 Fig. 4. Calibration of the detector in electron beam 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 0.88 0.90 0.92 0.94 0.96 0.98 1.00 D os e, re l.u n. Z (cм) Fig. 5. Depth dose distribution in absorber of phantom 2.2. The results of simulation of the dose distribution within the phantom absorber are shown in Fig. 5. It is seen, that the dose in the median plane of the phantom (z = 0.9 cm) corresponds to its average value over the volume of the absorber, measured by a calorimetric method. 2.3. The phantom and detector were irradiated with the electron beam having the parameters:  electron energy, MeV 9.8;  mean current, µA 760;  pulse rate, Hz 250;  pulse duration, µsec 3.5;  width of the beam scanning zone, cm 32;  scanning frequency, Hz 3. The results of measurements are given in Table 1. Table 1 Parameters of DD calibration in electron beam Dose rate, kGy/s DD current, µA DD sensitivity ×1010, C/Gy 2.13 0.77 3.61 1.08 0.36 3.34 0.54 0.17 3.09 3. X-RAY 3.1. For measurement of detector sensitivity in the bremsstrahlung field with end-point energy 10 MeV, a converter, cooled with water, was installed at the output of the LU-10 accelerator (Fig. 6). The converter design and the parameters of the radiation field are described in details in Ref. [10]. In particular, it was demonstrated that the electrons of a primary beam are almost com- pletely absorbed in the converter. So a practically "pure" X-ray flux affected the DD. The latter was stacked with the RP dosimeter and positioned at a distance of 40 cm downstream the converter. Following them, a free-air wide-aperture ionization chamber (WIC), which moni- tored the energy flux of X-ray, was placed at a distance of ~1.5 m. LU-10 +U DD C Н2О Iion 40cm RPe- Fig. 6. Calibration of DD in X-ray radiation The measurements were conducted at a different pulse repetition rate of the accelerator: 250, 125 and 62.5 Hz. The exposition changed in the range 20…60 min providing the absorbed dose values within an operating range of the RP dosimeters (Table 2). Table 2 Results of DD calibration (X-ray, 10 MeV) Pulse repetition rate, Hz Average dose rate, Gy/s DD current, nA DD sensi- tivity ×1010, C/Gy 62.5 2.97 0.8 2.75 125 5.72 1.8 3.14 250 13.32 4.0 3.0 ISSN 1562-6016. ВАНТ. 2014. №3(91) 164 3.2. Determination of detector sensitivity to the X- ray radiation with end-point energy 26 MeV was carried out at an accelerator EPOS. In this case, the bremsstrah- lung field was formed by interaction of a scanned elec- tron beam with a technological target [11]. In the meas- urements, the accelerator generated the beam with fol- lowing parameters: o electron energy, MeV 26; o average beam current, μA 420; o pulse duration, μsec 4.2; o pulse repetition rate, Hz 150. The average dose rate of bremsstrahlung, measured with the RP detectors, was 4.90 Gy/s. The DD current was 1.25 nA, that corresponds to the detector sensitivity of 2.6×10–10 C/Gy. 3.3. Measurement of the DD sensitivity to brems- strahlung with end-point energy 40 MeV was conducted at an accelerator LU-40 [12] (Fig. 7). A converter C consisted of four tantalum plates, each of 1 mm thick and 30 mm in diameter and divided with the gaps of 1 mm for cooling. Downstream the converter, a filter- absorber of electrons F (an aluminum cylinder having the diameter and thickness of 40 mm) was positioned followed by the DD detector and RP dosimeter. LU-40 e- C F FAl Ubias Ta RP DD Fig. 7. Scheme of DD calibration (X-ray, 40 MeV) The accelerator provided the beam with following parameters:  electron energy, MeV 40;  average beam current, μA 4.5;  beam pulsed current, mA 55  pulse width, μs 1.5;  pulse repetition rate, Hz 50. The measurements have shown, that the average bremsstrahlung dose rate in DD was 3.0 kGy/s and the DD current -0.9 nA. Thus, the detector sensitivity makes 3.0×10–10 C/Gy. 3.4. For an independent analysis of the measurement data, a computer simulation of the DD calibration con- ditions in the bremsstrahlung field was carried out. The developed simulation model included the description of a primary beam with the actual spatial and energy dis- tribution of electrons, its interaction with the converter and the influence of a secondary radiation produced on the target system. 20 30 40 50 60 70 0 2 4 6 8 10 12 14 Electron tnergy, MeV A bs or be d do se , J /µ A ⋅h ⋅g Fig. 8. Dependence of absorbed dose in DD and RP on electron beam energy Fig. 8 displays the calculated absorbed dose in DD and RP (solid lines) as well as the measured dose in the RP (dots). It is seen, that the calculated and measured data are in good agreement confirming the correctness of the DD calibration. This result was expected, taking into account the proximity of atomic numbers (~6) of sensitive volumes in DD and RP. 3.5. The results of evaluation of a specific sensitivity of DD to the electron and X-ray radiation are summa- rized in Table 3. These data show that the DD sensitivi- ty within an error of measurement (±15%) does not de- pend on the type and energy of radiation. The major sources of the error are those, connected with applica- tion of the Red Perspex dosimeters, as well as with de- termination of the detector average current using its pulse value. Table 3 Specific sensitivity of DD Radiation Energy, MeV Specific sensitivity ×1010, C/Gy·mm3 Electron 9.8 1.81 Photon 10.0 26.0 40.0 1.64 1.45 1.67 CONCLUSIONS The CVD-diamond detectors can be applied for technological dosimetry of electron and bremsstrahlung radiation in the range of dose rate 3…2500 Gy/s at the electron and photon energy 10…40 MeV with radiation durability not less than ~107 Gy. The average value of specific sensitivity, obtained in this work for an experimental detector, is twice as much as that determined at the DD calibration under low radi- ation dose rate [4]. The difference is most likely due to the specificities of the applied techniques for the CVD- diamond grows and, as a result, to the detector parame- ters obtained. REFERECES 1. S.F. Kozlov, E.A. Konorova, Y.A. Salikov, et al. Diamond Dosimeter for X-Ray and β-radiation // IEEE Trans. Nucl. Sci. 1977, v. 24, p. 235-237. 2. P. Bergonzo, A. Brambilla, D. Tomson, et al. CVDdiamond for Nuclear Detection Applications // NIM A. 2002, v. 476, p. 694-700. 3. M. Franklin, A. Fry, K.K. Gan, et al. Development of Diamond Radiation Detectors for SSC and LHC // NIM A. 1992, v. 315, p. 39-42. 4. G.A.P. Cirrone, G. Cuttone, S. Lo Nigro, et al. Do- simetric Characterization of CVD diamonds in Pho- ton, Electron and Proton Beams // Nucl. Phys. 2006, v. 150, p. 330-333. 5. L.N. Davydov, A.V. Rybka, A.A. Vierovkin, et al. Registration of High-Intensity Electron and X-ray Fields with Polycrystalline CVD-Diamond Detectors // Proc. of SPIE. 2012, 8507/N1-doi: 10.1117/ 12.929291. 6. ISO/ASTM 51276-12. Standard Practice for Use of a Polymethylmethacrylate Dosimetry System. 7. F. Salvat, J.M. Fernandez-Varea, I. Sempau. PENELOPE-2008 a Code System for Monte-Carlo ISSN 1562-6016. ВАНТ. 2014. №3(91) 165 Simulation of Electron and Photon Transport // OCED Nucl. Ener. Agency (Issy-les-Moulineaus) France. 2008. 8. V.N. Boriskin et al. Development of radiation tech- nology and tests in “Accelerator” Sc&Res Est NSC KIPT // Problems of Atomic Science and Technolo- gy. Ser. “Nuclear Physics Investigations” (50). 2008, №5, p. 150-154. 9. ISO/ASTM 51631:2011(E). Practice for Use of Cal- orimetric Dosimetry Systems for Electron Dose Measurements and Routine Dosimetry System. 10. V.I. Nikiforov, V.L. Uvarov. A Method for Analysis and Optimization of Electron Accelerator Exit De- vices under Breamsstrahlung Generation Mode // Problems of Atomic Science and Technology. Ser. “Nuclear Physics Investigations” (58). 2012, №3, p. 207-210. 11. A.N. Dovbnya, N.A. Dovbnya, V.I. Nikiforov, V.L. Uvarov. Electron Linac Based e, X-Facillity // Proc. of EPAC Edinburgh, Scotland, 2006, p. 1235- 1237. 12. N.I. Ayzatsky, V.I. Beloglazov, V.P. Bozhko, et al. Electron 100 MeV Linac Based Facility for Nuclear- Physics Experimental Investigation // Problems of Atomic Science and Technology. Ser. “Nuclear Physics Investigations”. (53). 2010, №2, p. 18-22. Article received 04.11.2013 КАЛИБРОВКА ДОЗИМЕТРОВ НА ОСНОВЕ CVD-АЛМАЗА В ПОЛЯХ ЭЛЕКТРОННОГО И ТОРМОЗНОГО ИЗЛУЧЕНИЙ БОЛЬШОЙ МОЩНОСТИ В.Е. Кутний, В.И. Никифоров, О.А. Опалев, Ю.В. Рогов, А.В. Рыбка, В.А. Шевченко, И.Н. Шляхов, В.Е. Стрельницкий, А.Э. Тенишев, В.Л. Уваров Описаны результаты исследования дозиметрических характеристик изготовленного в ННЦ ХФТИ опыт- ного образца детектора на основе поликристаллической алмазной пленки (CVD-алмаз). Разработаны мето- дики калибровки детекторов по мощности поглощенной дозы электронного и тормозного излучений. Усло- вия калибровки исследованы методом компьютерного моделирования. Для определения чувствительности детектора к электронному излучению его помещали внутри фантома из полистирола. При измерениях в поле тормозного излучения использовался дополнительный фильтр электронов. В каждом случае облучение де- тектора выполнялось совместно с дозиметрами Harwell Red Perspex 4034, по показаниям которых произво- дилась калибровка. Проведенные на ускорителях ЛУ-10, ЭПОС и ЛУ-40 ННЦ ХФТИ измерения показали, что значения чувствительности детектора для каждого вида излучения близки и согласуются с полученными ранее в условиях полей излучения низкой интенсивности. Учитывая высокую радиационную стойкость CVD-алмаза, это обеспечивает возможность его применения в технологической дозиметрии. КАЛІБРУВАННЯ ДОЗИМЕТРІВ НА ОСНОВІ CVD-АЛМАЗУ В ПОЛЯХ ЕЛЕКТРОННОГО ТА ГАЛЬМІВНОГО ВИПРОМІНЮВАНЬ ВЕЛИКОЇ ПОТУЖНОСТІ В.Є. Кутній, В.І. Нікіфоров, О.А. Опальов, Ю.В. Рогов, А.В. Рибка, В.А. Шевченко, І.М. Шляхов, В.Є. Стрельницький, А.Е. Тєнішев, В.Л. Уваров Описанo результати дослідження дозиметричних характеристик виготовленого в ННЦ ХФТІ дослідного зразка детектора на основі полікристалічної алмазної плівки (CVD-алмаз). Розроблено методики калібру- вання детекторів за потужністю поглинутої дози електронного та гальмівного випромінювань. Методом комп'ютерного моделювання досліджені умови калібрувань. Для визначення чутливості детектора до елект- ронного випромінювання його розміщували усередині стандартного фантома з полістиролу. При вимірю- ваннях у полі гальмівного випромінювання використовувався додатковий фільтр електронів. У кожному випадку опромінювання детектора виконувалося спільно з дозиметрами Harwell Red Perpex 4034, за пока- заннями яких проводилося калібрування. Проведені на прискорювачах ЛУ-10, ЕПОС та ЛУ-40 ННЦ ХФТІ вимірювання показали, що значення чутливості детектора для кожного виду випромінювання є близькі і узгоджуються з такими, що одержані раніше в умовах полів випромінювання низької інтенсивності. Зважа- ючи на значну радіаційну стійкість CVD-алмазу, це забезпечує можливість його застосування в технологіч- ній дозиметрії. INTRODUCTION 2. ELECTRON RADIATION 3. X-RAY CONCLUSIONS REFERECES
id nasplib_isofts_kiev_ua-123456789-80276
institution Digital Library of Periodicals of National Academy of Sciences of Ukraine
issn 1562-6016
language English
last_indexed 2025-12-07T15:25:52Z
publishDate 2014
publisher Національний науковий центр «Харківський фізико-технічний інститут» НАН України
record_format dspace
spelling Kutny, V.E.
Nikiforov, V.I.
Opalev, O.A.
Rogov, Yu.V.
Rybka, A.V.
Shevchenko, V.A.
Shlyakhov, I.N.
Strelnitsky, V.E.
Tenishev, A.Eh.
Uvarov, V.L.
2015-04-14T06:21:06Z
2015-04-14T06:21:06Z
2014
Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields / V.E. Kutny, V.I. Nikiforov, O.A. Opalev, Yu.V. Rogov, A.V. Rybka, V.A. Shevchenko, I.N. Shlyakhov, V.E. Strelnitsky, A.Eh. Tenishev, V.L. Uvarov // Вопросы атомной науки и техники. — 2014. — № 3. — С. 162-165. — Бібліогр.: 12 назв. — англ.
1562-6016
PACS: 07.57.Kp, 29.40.-n, 85.25.Pb, 81.05.ug
https://nasplib.isofts.kiev.ua/handle/123456789/80276
Results of a study of dosimetry characteristics of a prototype of the detector based on a polycrystalline diamond film (CVD-diamond) produced in NSC KIPT are summarized. The techniques of the detector calibration against the electron and X-ray radiation dose rate are developed. The conditions of calibration were studied by means of computer simulation. For determination of detector sensitivity to electron radiation, it was placed inside of a standard polystyrene phantom. An additional filter of electrons was used at measurement in the X-ray field. Every time detector irradiation was carried out together with the Harwell Red Perpex 4034 dosimeters, which registrations provided the calibration data. The measurements, executed at the LU-10, EPOS and LU-40 linacs of NSC KIPT, have demonstrated, that the values of detector sensitivity against each type of radiation are close and conform to ones obtained earlier for the low-intensity radiation fields. Considering significant radiation durability of the CVD-diamond, it enables its application in technological dosimetry.
Описаны результаты исследования дозиметрических характеристик изготовленного в ННЦ ХФТИ опытного образца детектора на основе поликристаллической алмазной пленки (CVD-алмаз). Разработаны методики калибровки детекторов по мощности поглощенной дозы электронного и тормозного излучений. Условия калибровки исследованы методом компьютерного моделирования. Для определения чувствительности детектора к электронному излучению его помещали внутри фантома из полистирола. При измерениях в поле тормозного излучения использовался дополнительный фильтр электронов. В каждом случае облучение детектора выполнялось совместно с дозиметрами Harwell Red Perspex 4034, по показаниям которых производилась калибровка. Проведенные на ускорителях ЛУ-10, ЭПОС и ЛУ-40 ННЦ ХФТИ измерения показали, что значения чувствительности детектора для каждого вида излучения близки и согласуются с полученными ранее в условиях полей излучения низкой интенсивности. Учитывая высокую радиационную стойкость CVD-алмаза, это обеспечивает возможность его применения в технологической дозиметрии.
Описанo результати дослідження дозиметричних характеристик виготовленого в ННЦ ХФТІ дослідного зразка детектора на основі полікристалічної алмазної плівки (CVD-алмаз). Розроблено методики калібрування детекторів за потужністю поглинутої дози електронного та гальмівного випромінювань. Методом комп'ютерного моделювання досліджені умови калібрувань. Для визначення чутливості детектора до електронного випромінювання його розміщували усередині стандартного фантома з полістиролу. При вимірюваннях у полі гальмівного випромінювання використовувався додатковий фільтр електронів. У кожному випадку опромінювання детектора виконувалося спільно з дозиметрами Harwell Red Perpex 4034, за показаннями яких проводилося калібрування. Проведені на прискорювачах ЛУ-10, ЕПОС та ЛУ-40 ННЦ ХФТІ вимірювання показали, що значення чутливості детектора для кожного виду випромінювання є близькі і узгоджуються з такими, що одержані раніше в умовах полів випромінювання низької інтенсивності. Зважаючи на значну радіаційну стійкість CVD-алмазу, це забезпечує можливість його застосування в технологічній дозиметрії.
en
Національний науковий центр «Харківський фізико-технічний інститут» НАН України
Вопросы атомной науки и техники
Применение ускорителей в радиационных технологиях
Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields
Калибровка дозиметров на основе CVD-алмаза в полях электронного и тормозного излучений большой мощности
Калібрування дозиметрів на основі CVD-алмазу в полях електронного та гальмівного випромінювань великої потужності
Article
published earlier
spellingShingle Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields
Kutny, V.E.
Nikiforov, V.I.
Opalev, O.A.
Rogov, Yu.V.
Rybka, A.V.
Shevchenko, V.A.
Shlyakhov, I.N.
Strelnitsky, V.E.
Tenishev, A.Eh.
Uvarov, V.L.
Применение ускорителей в радиационных технологиях
title Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields
title_alt Калибровка дозиметров на основе CVD-алмаза в полях электронного и тормозного излучений большой мощности
Калібрування дозиметрів на основі CVD-алмазу в полях електронного та гальмівного випромінювань великої потужності
title_full Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields
title_fullStr Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields
title_full_unstemmed Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields
title_short Calibration of CVD-diamond based dosimeter in high-power electron and X-ray radiation fields
title_sort calibration of cvd-diamond based dosimeter in high-power electron and x-ray radiation fields
topic Применение ускорителей в радиационных технологиях
topic_facet Применение ускорителей в радиационных технологиях
url https://nasplib.isofts.kiev.ua/handle/123456789/80276
work_keys_str_mv AT kutnyve calibrationofcvddiamondbaseddosimeterinhighpowerelectronandxrayradiationfields
AT nikiforovvi calibrationofcvddiamondbaseddosimeterinhighpowerelectronandxrayradiationfields
AT opalevoa calibrationofcvddiamondbaseddosimeterinhighpowerelectronandxrayradiationfields
AT rogovyuv calibrationofcvddiamondbaseddosimeterinhighpowerelectronandxrayradiationfields
AT rybkaav calibrationofcvddiamondbaseddosimeterinhighpowerelectronandxrayradiationfields
AT shevchenkova calibrationofcvddiamondbaseddosimeterinhighpowerelectronandxrayradiationfields
AT shlyakhovin calibrationofcvddiamondbaseddosimeterinhighpowerelectronandxrayradiationfields
AT strelnitskyve calibrationofcvddiamondbaseddosimeterinhighpowerelectronandxrayradiationfields
AT tenishevaeh calibrationofcvddiamondbaseddosimeterinhighpowerelectronandxrayradiationfields
AT uvarovvl calibrationofcvddiamondbaseddosimeterinhighpowerelectronandxrayradiationfields
AT kutnyve kalibrovkadozimetrovnaosnovecvdalmazavpolâhélektronnogoitormoznogoizlučeniibolʹšoimoŝnosti
AT nikiforovvi kalibrovkadozimetrovnaosnovecvdalmazavpolâhélektronnogoitormoznogoizlučeniibolʹšoimoŝnosti
AT opalevoa kalibrovkadozimetrovnaosnovecvdalmazavpolâhélektronnogoitormoznogoizlučeniibolʹšoimoŝnosti
AT rogovyuv kalibrovkadozimetrovnaosnovecvdalmazavpolâhélektronnogoitormoznogoizlučeniibolʹšoimoŝnosti
AT rybkaav kalibrovkadozimetrovnaosnovecvdalmazavpolâhélektronnogoitormoznogoizlučeniibolʹšoimoŝnosti
AT shevchenkova kalibrovkadozimetrovnaosnovecvdalmazavpolâhélektronnogoitormoznogoizlučeniibolʹšoimoŝnosti
AT shlyakhovin kalibrovkadozimetrovnaosnovecvdalmazavpolâhélektronnogoitormoznogoizlučeniibolʹšoimoŝnosti
AT strelnitskyve kalibrovkadozimetrovnaosnovecvdalmazavpolâhélektronnogoitormoznogoizlučeniibolʹšoimoŝnosti
AT tenishevaeh kalibrovkadozimetrovnaosnovecvdalmazavpolâhélektronnogoitormoznogoizlučeniibolʹšoimoŝnosti
AT uvarovvl kalibrovkadozimetrovnaosnovecvdalmazavpolâhélektronnogoitormoznogoizlučeniibolʹšoimoŝnosti
AT kutnyve kalíbruvannâdozimetrívnaosnovícvdalmazuvpolâhelektronnogotagalʹmívnogovipromínûvanʹvelikoípotužností
AT nikiforovvi kalíbruvannâdozimetrívnaosnovícvdalmazuvpolâhelektronnogotagalʹmívnogovipromínûvanʹvelikoípotužností
AT opalevoa kalíbruvannâdozimetrívnaosnovícvdalmazuvpolâhelektronnogotagalʹmívnogovipromínûvanʹvelikoípotužností
AT rogovyuv kalíbruvannâdozimetrívnaosnovícvdalmazuvpolâhelektronnogotagalʹmívnogovipromínûvanʹvelikoípotužností
AT rybkaav kalíbruvannâdozimetrívnaosnovícvdalmazuvpolâhelektronnogotagalʹmívnogovipromínûvanʹvelikoípotužností
AT shevchenkova kalíbruvannâdozimetrívnaosnovícvdalmazuvpolâhelektronnogotagalʹmívnogovipromínûvanʹvelikoípotužností
AT shlyakhovin kalíbruvannâdozimetrívnaosnovícvdalmazuvpolâhelektronnogotagalʹmívnogovipromínûvanʹvelikoípotužností
AT strelnitskyve kalíbruvannâdozimetrívnaosnovícvdalmazuvpolâhelektronnogotagalʹmívnogovipromínûvanʹvelikoípotužností
AT tenishevaeh kalíbruvannâdozimetrívnaosnovícvdalmazuvpolâhelektronnogotagalʹmívnogovipromínûvanʹvelikoípotužností
AT uvarovvl kalíbruvannâdozimetrívnaosnovícvdalmazuvpolâhelektronnogotagalʹmívnogovipromínûvanʹvelikoípotužností