Влияние нестабильности потока излучения ртутных ламп на калибровку приборов

The instability of the radiation flux of gas-discharge mercury lamps used in the calibration of optoelectronic instruments has been investigated. A methodology for accounting for this instability is presented, which makes it possible to reduce the calibration error of UV radiometers from ±10.2% to...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Datum:2006
Hauptverfasser: Doktorovich, I. V., Fodchuk, I. M., Butenko, V. K., Godovanyuk, V. N., Ryukhtin, V. V., Yuryev, V. G.
Format: Artikel
Sprache:Ukrainisch
Veröffentlicht: PE "Politekhperiodika", Book and Journal Publishers 2006
Schlagworte:
Online Zugang:https://www.tkea.com.ua/index.php/journal/article/view/TKEA2006.1.54
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Назва журналу:Technology and design in electronic equipment

Institution

Technology and design in electronic equipment
Beschreibung
Zusammenfassung:The instability of the radiation flux of gas-discharge mercury lamps used in the calibration of optoelectronic instruments has been investigated. A methodology for accounting for this instability is presented, which makes it possible to reduce the calibration error of UV radiometers from ±10.2% to ±8.5%. It is not recommended to use mercury lamps for measuring the instability of photodetector sensitivity.