A HIGH SENSITIVE MICROWAVE MEASURING DEVICE OF THE MOISTURE CONTENT IN THE NON-POLAR DIELECTRIC LIQUIDS BASED ON AN INHOMOGENEOUS STEP COAXIAL RESONATOR
Purpose. Objective is to create a moisture meter for non-polar liquid dielectrics with low volumetric moisture content of more than 10‑3 %. Methodology. Moisture measuring is based on dielcometric method. It is implemented as a resonant method of determining a capacitance measuring transducer. Measu...
Saved in:
| Date: | 2016 |
|---|---|
| Main Authors: | Rudakov, V. V., Korobko, A. A. |
| Format: | Article |
| Language: | English Russian |
| Published: |
National Technical University "Kharkiv Polytechnic Institute" and Аnatolii Pidhornyi Institute of Power Machines and Systems of NAS of Ukraine
2016
|
| Subjects: | |
| Online Access: | http://eie.khpi.edu.ua/article/view/2074-272X.2016.5.08 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| Journal Title: | Electrical Engineering & Electromechanics |
Institution
Electrical Engineering & ElectromechanicsSimilar Items
-
MULTIFREQUENCY ALGORITHMS FOR DETERMINING THE MOISTURE CONTENT OF LIQUID EMULSIONS BY THE METHOD OF RESONANCE DIELCOMETRY
by: Korobko, A. A.
Published: (2017) -
WAYS TO REDUCE ERRORS IN MEASURING THE DIELECTRIC CONSTANT OF WEAKLY ABSORBING DIELECTRICS IN THE MILLIMETER AND SUBMILLIMETER (TERAHERTZ) WAVELENGTH RANGES BY THE SURFACE PLASMON RESONANCE METHOD
by: Gavrikov, V. K., et al.
Published: (2024) -
RELAXATIONS LOSSES IN POLYETHYLENE INSULATION OF COAXIAL CABLE STRUCTURE DURING AGING IN HIGH HUMIDITY CONDITIONS
by: Bezprozvannych, G. V., et al.
Published: (2016) -
ALGORITHMS OF POLARIZATION PARAMETERS DETERMINATION OF PULSAR RADIO EMISSION
by: Ulyanov, O. M., et al.
Published: (2014) -
THE METHOD OF DETERMINING THE DIELECTRIC RELATIVE PERMITTIVITY IN THE MM AND SUBMM WAVELENGTH RANGES BASED ON THE MEASURING OF THE PLASMON-POLARITONE RESONANCE PARAMETERS
by: Lytvynenko, L. N., et al.
Published: (2020)