Расчет коэффициента преобразования кондуктометрического датчика биосенсора
A five-component model for calculating the transformation factor of planar comb electrodes in a conductimetric biosensor is proposed. The existing three-component model is extended by two additional terms that account for the electric field at electrode corners and electrode thickness. It is also su...
Saved in:
| Date: | 2007 |
|---|---|
| Main Authors: | , |
| Format: | Article |
| Language: | Ukrainian |
| Published: |
PE "Politekhperiodika", Book and Journal Publishers
2007
|
| Subjects: | |
| Online Access: | https://www.tkea.com.ua/index.php/journal/article/view/TKEA2007.5.35 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| Journal Title: | Technology and design in electronic equipment |
Institution
Technology and design in electronic equipment| Summary: | A five-component model for calculating the transformation factor of planar comb electrodes in a conductimetric biosensor is proposed. The existing three-component model is extended by two additional terms that account for the electric field at electrode corners and electrode thickness. It is also suggested to round the electrode corners. Graphical results of the conductimetric sensor constant calculations are presented. |
|---|