Влияние нестабильности потока излучения ртутных ламп на калибровку приборов
The instability of the radiation flux of gas-discharge mercury lamps used in the calibration of optoelectronic instruments has been investigated. A methodology for accounting for this instability is presented, which makes it possible to reduce the calibration error of UV radiometers from ±10.2% to...
Saved in:
| Date: | 2006 |
|---|---|
| Main Authors: | , , , , , |
| Format: | Article |
| Language: | Ukrainian |
| Published: |
PE "Politekhperiodika", Book and Journal Publishers
2006
|
| Subjects: | |
| Online Access: | https://www.tkea.com.ua/index.php/journal/article/view/TKEA2006.1.54 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| Journal Title: | Technology and design in electronic equipment |
Institution
Technology and design in electronic equipment| Summary: | The instability of the radiation flux of gas-discharge mercury lamps used in the calibration of optoelectronic instruments has been investigated. A methodology for accounting for this instability is presented, which makes it possible to reduce the calibration error of UV radiometers from ±10.2% to ±8.5%. It is not recommended to use mercury lamps for measuring the instability of photodetector sensitivity.
|
|---|